Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65672

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 28, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Alphatec Spine, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Illico Ti Cannulated Polyaxial Screw 7.5mm X 40mm. The Illico MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development.

Z-2086-2013
Recall number
Z-2086-2013
Initiated
June 28, 2013
Classification
Class II
Status
Terminated
Recalling firm
Alphatec Spine, Inc.
Quantity
333 screws

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that when the screw is mated to a polyaxial screwdriver, the guide wire may not pass through the screwdriver.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that when the screw is mated to a polyaxial screwdriver, the guide wire may not pass through the screwdriver.

Code information

Part Number 73875-40. Lots Affected: 647289, 648923, 649244, 649556, 649813, 650040, 650262, 650589.

Distribution pattern

Worldwide distribution: US (nationwide) including states of:US states: AZ, CA, FL, GA, IL, MD, NJ, NY, OK, PA, TN, TX,and WA ; and countries of: Spain, France, Japan. International: Spain, France, Japan.