device · product 1 of 1
Brilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/N 728231, Philips Healthcare, Cleveland, OH 44143. Intended to produce cross-sectional images of the body by computer reconstruction.
- Recall number
- Z-2210-2013
- Initiated
- July 05, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Quantity
- M/N: 728256, 4 units; M/N: 728246, 89 units; M/N: 728235, 9 units & M/N: 728231, 207 units
App-derived interpretation
The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.
Official device-enrichment evidence · Sourced
Component design/selection
Inspect official wording and provenance
Reason for recall
The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.
Code information
Model #, 728231, Serial #: 4021, 9535, 9544, 9587, 9627, 9701, 9707, 9713, 9759, 9875, 9887, 9917, 9925, 9931, 9980, 9996, 10009, 10011, 10012, 10015, 10028, 10036, 10042, 10045, 10050, 10052, 10062, 10063, 10064, 10077, 10079, 10081, 10096, 10099, 10102, 10111, 10114, 10120, 10122, 10135, 10136, 10137, 10139, 10145, 10148, 10149, 10151, 10152, 10153, 10154, 10162, 10165, 10166, 10170, 10171, 10172, 10173, 10188, 10189, 10191, 10196, 10198, 10203, 10214, 10222, 10227, 10228, 10229, 10230, 10235, 10238, 10242, 10243, 10247, 10259, 10260, 10261, 10263, 10268, 10273, 10274, 10278, 10288, 10291, 10293, 10294, 10297, 10298, 10301, 10302, 10307, 10312, 10313, 10316, 10317, 10318, 10319, 10320, 10327, 10336, 10337, 10341, 10347, 10349, 10350, 10351, 10352, 10353, 10362, 10365, 10376, 10380, 10384, 10387, 10393, 10400, 10411, 10412, 10414, 10431, 10438, 10439, 10445, 10446, 10452, 10453, 10459, 10461, 10468, 10470, 10471, 10476, 10482, 10486, 10490, 10491, 10503, 10511, 10513, 10519, 10535, 10536, 10540, 10544, 10546, 10551, 10554, 10561, 10563, 10568, 10569, 10574, 10582, 10600, 10617, 10628, 10631, 10635, 10646, 10647, 10651, 10655, 10659, 10673, 10684, 10686, 10698, 10714, 10733, 10734, 29002, 29014, 29016, 29027, 90054, 90099, 95022, 95046, 95073, 95103, 95107, 95109, 95114, 95132, 95133, 95181, 95209, 95268, 95295, 95305, 95371, 95453, 95473, 95563, 95660, 95704, 95706, 95707, 95712, 95715, 95721, 95727, 95729, 95769, 95778, 90199 & 95162. Model #, 728235, Serial #: 9034, 9190, 9233, 9234, 9261, 29049, 29057, 29064 & 29149. Model #, 728246, Serial #: 3150, 3163, 4051, 5547, 5585, 5623, 5655, 5666, 5670, 5715, 5724, 5747, 5773, 5776, 5791, 5803, 5813, 5836, 5855, 5862, 5868, 5876, 5877, 5897, 5911, 5916, 5935, 5969, 5971, 5974, 5997, 6001, 6006, 6008, 6028, 6087, 6102, 6117, 6120, 6121, 6126, 6170, 6195, 6200, 6202, 6203, 6238, 6243, 6281, 6287, 6306, 6318, 6324, 6346, 6352, 6387, 6402, 6405, 6444, 6484, 6491, 6494, 6495, 6496, 6503, 6517, 6519, 6523, 6556, 6559, 6609, 6610, 6611, 6612, 6623, 6624, 6640, 6642, 50115, 50147, 50166, 50180, 50189, 50193, 50195, 50197, 50198, 50204 & 50043. Model #, 728256, Serial #: 3490, 3500, 3679 & 3693.
Distribution pattern
Worldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, DE, FL, GA, IN, KY, MA, MD, MI, MN, MS, NC, NM, NV, NY, OH, OK, PA, SC, TN, TX, VT, WI & WV, and the countries of Australia, Austria, Bosnia & Herzegovina, Chile, China, Croatia, Denmark, Equatorial Guinea, France, Germany, Greenland, Iceland, Indonesia, Israel, Japan, Kazakhstan, Netherlands, Nigeria, Norway, Poland, Romania, South Africa, South Korea, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom & Vietnam.