Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65787

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 30, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Craniomaxillofacial Division

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Stryker Part Numbers 79-43903 (3cc), 79-43905 (5cc), 79-43910 (10 cc), 79-43915 (15cc), 397003 (3cc), 397005 (5cc), 397010 (10 cc), 397015 (15 cc) HydroSet Injectable HA Bone Substitute Sterile R/EO, nonpyrogenic, Product of Ireland Stryker Injectable Cement is a self-setting, calcium phosphate cement intended for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects as well as in the augmentation or restoration of bony contour in the craniofacial skeleton. The powder is mixed with the sodium phosphate/PVP solution provided in the kit and loaded into the empty cement delivery syringe to be applied directly to the site for void filling or augmentation.

Z-2054-2013
Recall number
Z-2054-2013
Initiated
July 30, 2013
Classification
Class II
Status
Terminated
Quantity
12,719

App-derived interpretation

Foreign material or chemical contamination reason.foreign_material_chemical_contamination · v1.0.0
foreign body

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Stryker received reports of the desiccant bag in the kit leaking silica gel granulate. This could increase exposure to anesthesia during surgical delay up to 60 minutes or inadvertent implantation of silica gel granulate which could cause foreign body reaction including inflammation and might lead to the need for revision surgery.

Code information

Part 79-43903, Lot: IC01116, IC01130, IC01204, IC01233, IC01238 Part 79-43905 Lot: IC01084, IC01085, IC01086, IC01093, IC01123, IC01150, IC01166, IC01167, IC01168, IC01180, IC01188, IC01189, IC01190, IC01198, IC01199, IC01200, IC01213, IC01214, IC01229, IC01230, IC01239, IC01240, IC01245, IC01246, IC01270 Part 79-43910 Lot: IC01087, IC01090, IC01091, IC01153, IC01154, IC01163, IC01164, IC01165, IC01179, IC01183, IC01184, IC01201, IC01202, IC01203, IC01231, IC01243, IC01244, IC01248, IC01249, IC01261, IC01264, Part 79-43015 Lot: IC01100, IC01114, IC01161, IC01162, IC01182, IC01211, IC01212, IC01227, IC01228, IC01241, IC01242, IC01252, IC01253, IC01254, IC01267 Part 397003 Lot: IC01092, IC01101, IC01102, IC01103, IC01104, IC01105, IC01106, IC01107, IC01115, IC01127, IC01131, IC01205, IC01220, IC01232, IC01234 Part 397005 Lot: IC01088, IC01094, IC01095, IC01096, IC01097, IC01098, IC01110, IC01111, IC01118, IC01119, IC01120, IC01132, IC01133, IC01134, IC01135, IC01136, IC01137, IC01147, IC01148, IC01149, IC01158, IC01159, IC01160, IC01171, IC01172, IC01181, IC01193, IC01194, IC01195, IC01196, IC01223, IC01224, IC01247Part 79-43910 Lot: IC01087, IC01090, IC01091, IC01153, IC01154 Part 397010 Lot: IC01099, IC01108, IC01109, IC01112, IC01121, IC01138, IC01139, IC01140, IC01141, IC01151, IC01152, IC01155, IC01156, IC01157, IC01173, IC01174, IC01191, IC01192, IC01197, IC01210, IC01216, IC01219, IC01225, IC01226, IC01250, IC01251, IC01263, IC01266, IC01301 Part 397015 Lot: IC01113, IC01122, IC01124, IC01125, IC01142, IC01143, IC01144, IC01145, IC01146, IC01169, IC01170, IC01177, IC01178, IC01185, IC01186, IC01187, IC01206, IC01207, IC01208, IC01209, IC01217, IC01218, IC01221, IC01222, IC01265

Distribution pattern

Worldwide Distribution - USA (nationwide) including the countries through Stryker Distribution: Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Ireland, Italy, Latvia, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Spain, Sweden, Switzerland, United Kingdom, Argentina, Australia, Brazil, Jordan, Kingdom of Saudi Arabia, Kuwait, Latin America (others), New Zealand, Russia, Singapore, South Africa, and United Arabian Emirates.