device · product 1 of 7
ACL TOP (Base); PN 0000280000 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.
- Recall number
- Z-2244-2013
- Initiated
- July 19, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Instrumentation Laboratory Co.
- Quantity
- 929
App-derived interpretation
Potential carryover issue that can can cause shortened APTT clotting times on the instrument.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Potential carryover issue that can can cause shortened APTT clotting times on the instrument.
Code information
Nov 2003 - Nov 2011, Serial Numbers 02110101 through 11111174
Distribution pattern
Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. Internationally to Canada, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Brunei, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dubai, Finland, France, Germany, Great Britian, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Luthuania, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uruguay, Venezuela, and Viet Nam.