Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66392

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 24, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Baxter Healthcare Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

SIGMA SPECTRUM Infusion Pump with Master Drug Library component (SHARP LCD screen, 55074-2 revision B) Product Usage: Intended to be used as a visual aid during the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.

Z-0189-2014
Recall number
Z-0189-2014
Initiated
September 24, 2013
Classification
Class II
Status
Terminated
Recalling firm
Baxter Healthcare Corp.
Quantity
346 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.

Code information

Product Code: 35700BAX & 35700ABB; Serial Numbers: 433713, 435563, 704056, 707290, 715315, 724216, 740127, 752318, 752447, 753675, 754947, 759766, 765934, 770034, 772017, 773470, 776507, 781016, 781969, 783352, 783454, 787567, 789720, 791225, 794828, 794990, 797185, 804798, 808393, 808764, 810971, 812076, 818718, 821213, 824067, 824198, 839143, 850028, 850577, 852007, 852431, 855450, 860201, 864799, 876765, 877548, 877944, 878637, 879810, 879886, 880022, 884707, 884778, 885565, 889416, 889814, 890949, 890987, 894957, 895413, 896662, 896912, 897576, 900401, 904778, 913511, 915035, 915981, 916173, 916948, 919159, 921162, 921937, 922515, 923861, 945583, 953011, 953360, 953995, 955881, 959847, 967694, 978892, 981692, 986022, 997333, 998512, 1002325, 1018432, 1018436, 1018737

Distribution pattern

Worldwide Distribution - USA Nationwide in the states of: AR, AZ, CA, CO, CT, FL, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NY, OH, OK, OR, PA, TN, UT, VA, WA and WV and the country Canada.