Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66410

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 27, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
S.A.E. AFIKIM

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Breeze C Scooter is a motorized electric scooter for elderly people. Brand name Breeze, model C. It is intended to provide increased mobility for elderly and/or disabled person and it is also an aid to independent living. Products can be identified by their color; the recalled products are either metallic blue or metallic silver. The products have two rear wheels and have either one or two wheels in the front. The scooters have an adjustable swivel seat and a front storage basket; some scooters may be fitted with an optional larger rear basket.

Z-0115-2014
Recall number
Z-0115-2014
Initiated
September 27, 2013
Classification
Class II
Status
Terminated
Recalling firm
S.A.E. AFIKIM
Quantity
69 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Breeze C Scooter, motorized 3-wheeled electric scooter for elderly people, is recalled because it may become unresponsive during operation due to an electrical contact which may fault due to extensive wear, which potentially could result in the driver not being able to stop the scooter during operation except by turning off the ignition key.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Breeze C Scooter, motorized 3-wheeled electric scooter for elderly people, is recalled because it may become unresponsive during operation due to an electrical contact which may fault due to extensive wear, which potentially could result in the driver not being able to stop the scooter during operation except by turning off the ignition key.

Code information

Serial Numbers: 10390024, 10390025, 10390029, 10390030, 10390034, 10390037, 10490100, 10490108, 10490111, 10490113, 10490119, 10490121, 10490124, 11390058, 11390060, 11390061, 11390069, 11390074, 11390095, 11390101, 11490111, 11490147, 11490176, 11490189, 11490277, 11490304, 11490305, 11490425, 11490426, 11490433, 11490435, 11490445, 12390096, 12390097, 12390098, 12390099, 12390101, 12390102, 12390103, 12390104, 12390105, 12390106, 12390107, 12390117, 12390120, 12490169, 12490170, 12490172, 12490173, 12490174, 12490175, 12490176, 12490177, 12490178, 12490179, 12490180, 12490181, 12490182, 12490183, 12490184, 12490185, 12490186, 12490187, 12490188, 12490189, 12490190, 12490191, 12490192, and 12490193.

Distribution pattern

Nationwide Distribution: US including the states of California, Florida, Georgia, Illinois, Maine, Maryland, Massachusetts, Michigan, Minnesota, New Jersey, New York, Ohio, Oklahoma, Pennsylvania, Texas, Virginia, and Wisconsin.