Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66420

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 31, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Synthes USA HQ, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Synthes Tensioning Device The Synthes Osteogenesis System, MR Conditional is intended for fracture fixation (open and closed), pseudoarthrosis or non-unions of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities and correction of segmental bony or soft tissue defects in adult and pediatric patients.

Z-0388-2014
Recall number
Z-0388-2014
Initiated
July 31, 2013
Classification
Class II
Status
Terminated
Recalling firm
Synthes USA HQ, Inc.
Quantity
369

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
All lots of the Synthes Tensioning Device were recalled due to the potential for the reaction fork to come loose from the main body of the device and not allow the wire to release, nor the device to apply tension.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

All lots of the Synthes Tensioning Device were recalled due to the potential for the reaction fork to come loose from the main body of the device and not allow the wire to release, nor the device to apply tension.

Code information

all lots of model number 03.311.001: Part # Synthes lot # 03.311.001 1934765 03.311.001 1934766 03.311.001 1993844 03.311.001 3036688 03.311.001 3398576 03.311.001 3776484 03.311.001 5100597 03.311.001 5106354 03.311.001 5160872 03.311.001 5160875 03.311.001 5174253 03.311.001 5174253 03.311.001 5174254 03.311.001 5181000 03.311.001 5181001 03.311.001 5196282 03.311.001 5196283 03.311.001 5197081 03.311.001 5241942 03.311.001 5241943 03.311.001 5336723 03.311.001 5363637 03.311.001 5396016 03.311.001 5411479 03.311.001 5555521

Distribution pattern

Nationwide Distribution.