Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66448

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 10, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Hospira Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Hospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL) Canister, ATS, 2000, with Valve 1) List Numbers 43449-01 and 43449-11; LIST NO. 43449 RECEPTAL CANISTER; CANISTER FOR REUSE, THROW AWAY LINER ONLY; DO NOT FILL 900 ABOVE LINE 2) List Numbers 43423-01 and 43423-21; LIST NO. 43423; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; 1400. 3) List Numbers 43445-01 and 43445-11; LIST NO. 43445; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; DO NOT FILL 1900 ABOVE THIS LINE. 4) List Number 43496-01.

Z-2593-2016
Recall number
Z-2593-2016
Initiated
September 10, 2013
Classification
Class II
Status
Terminated
Recalling firm
Hospira Inc.
Quantity
67,410,861 liners and 67,410,861 lids/canisters in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.

Code information

43423-04-01 43423-04-11 43423-04-21 43445-04-01 43445-04-11 43449-04-01 43449-04-11 43496-04-01

Distribution pattern

*** US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY), District of Columbia, Puerto Rico; *** FOREIGN: Australia, Canada, Costa Rica, Singapore, South Africa

device · product 2 of 2

Hospira RECEPTAL LINERS: 1L(1000mL), 1.5L (1500mL), 2L (2000mL) 1) List Number 43056-01; 2) List Number 43023-01; 3) List Number 43025-01; 4) List Number 43025-11; 5) List Number 43025-21; 6) List Number 43025-25; 7) List Number 43038-05; 8) List Number 43043-01; 9) List Number 43044-01; 10) List Number 43044-11; 11) List Number 43046-01; 12) List Number 43047-01; 13) List Number 43041-01; 14) List Number 43024-01; 15) List Number 43027-01; 16) List Number 43027-14; 17) List Number 43001-01; 18) List Number 43064-01; 19) List Number 43046-05; 20) List Number 43042-11; 21) List Number 43042-01; 22) List Number 43044-05; 23) List Number 43044-15

Z-2594-2016
Recall number
Z-2594-2016
Initiated
September 10, 2013
Classification
Class II
Status
Terminated
Recalling firm
Hospira Inc.
Quantity
67,410,861 liners and 67,410,861 lids/canisters in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.

Code information

43001-04-01 43023-04-01 43024-04-01 43024-04-62 43025-04-01 43025-04-11 43025-04-21 43025-04-25 43027-04-01 43027-04-05 43027-04-11 43027-04-14 43027-04-17 43031-04-02 43032-04-05 43034-04-06 43036-04-04 43037-04-02 43038-04-04 43038-04-05 43041-04-01 43042-04-01 43042-04-11 43043-04-01 43044-04-01 43044-04-05 43044-04-11 43044-04-15 43044-04-51 43046-04-01 43046-04-05 43047-04-01 43056-04-01 43061-04-01 43062-04-01 43063-04-01 43064-04-01 0L212-01 0L213-01

Distribution pattern

*** US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY), District of Columbia, Puerto Rico; *** FOREIGN: Australia, Canada, Costa Rica, Singapore, South Africa