Exact FDA event identifier

Event 66467

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 08, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ortho-Clinical Diagnostics

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

enGen Track System with enGen Select v5.0 with autoverification enabled Product The VITROS 5,1 FS Chemistry System with enGenTM Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.

Z-0941-2014
Recall number
Z-0941-2014
Initiated
October 08, 2013
Classification
Class II
Status
Terminated
Quantity
8 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ortho Clinical Diagnostics issued an Urgent Product Correction Notification involving the enGen Select V5.0 & enGen Custom v3.2.2 due to an anomaly that affects the autoverification feature and custom rules that depend upon data elements used by the referenced enGen Systems.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ortho Clinical Diagnostics issued an Urgent Product Correction Notification involving the enGen Select V5.0 & enGen Custom v3.2.2 due to an anomaly that affects the autoverification feature and custom rules that depend upon data elements used by the referenced enGen Systems.

Code information

Autoverification and Custom Rules for enGen Laboratory Automation Systems using enGenTM Select v5.0

Distribution pattern

Worldwide Distribution - USA Nationwide in the states and countries of: AZ, FL, WV, TX, PA, DE, MI, OH, TN, IL, NC, Chile, India, Italy, Portugal, Spain, Sweden, UK

device · product 2 of 2

enGen Track System with enGen Custom v3.2.2 (& below) Product The VITROS 5,1 FS Chemistry System with enGenTM Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.

Z-0942-2014
Recall number
Z-0942-2014
Initiated
October 08, 2013
Classification
Class II
Status
Terminated
Quantity
20 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ortho Clinical Diagnostics issued an Urgent Product Correction Notification involving the enGen Select V5.0 & enGen Custom v3.2.2 due to an anomaly that affects the autoverification feature and custom rules that depend upon data elements used by the referenced enGen Systems.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ortho Clinical Diagnostics issued an Urgent Product Correction Notification involving the enGen Select V5.0 & enGen Custom v3.2.2 due to an anomaly that affects the autoverification feature and custom rules that depend upon data elements used by the referenced enGen Systems.

Code information

Custom v3.2.2 & Below Configuration Files

Distribution pattern

Worldwide Distribution - USA Nationwide in the states and countries of: AZ, FL, WV, TX, PA, DE, MI, OH, TN, IL, NC, Chile, India, Italy, Portugal, Spain, Sweden, UK