Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66472

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 03, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems (Cleveland) Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

BrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT), one unit per box. BrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with an attenuation device consisting of flat panel x-ray imaging components. BrightView XCT produces non-attenuation corrected SPECT images and attenuation corrected SPECT images with x-ray transmission data that may also be used for scatter correction. The nuclear medicine images and the XCT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide anatomical localization of the nuclear medicine data.

Z-0089-2014
Recall number
Z-0089-2014
Initiated
September 03, 2013
Classification
Class II
Status
Terminated
Quantity
371 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Philips received reports from the field that the Flat Panel Detector (FPD) does not always remain securely latched in the stowed position.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Philips received reports from the field that the Flat Panel Detector (FPD) does not always remain securely latched in the stowed position.

Code information

The following model and serial numbers are affected by this recall: Model Number: 882454 Serial Number(s): 11000021, 11000082, and 11000111. Model Number: 882482 Serial Number(s): 6000002, 6000003, 6000004A, 6000005, 6000010, 6000011, 6000012, 6000013, 6000015, 6000016, 6000017, 6000018, 6000019, 6000020, 6000022, 6000023, 6000024, 6000025, 6000026, 6000027, 6000028, 6000029, 6000030, 6000031, 6000033, 6000036, 6000037, 6000038,6000040,6000041, 60000042, 6000043, 6000044, 6000045, 6000047, 6000048, 6000049, 6000050, 6000051, 6000052, 6000053, 6000054, 6000055, 6000056, 6000058, 6000059, 6000060, 6000061, 6000062, 6000063, 6000064, 6000065, 6000066, 6000067, 6000069, 6000070, 6000071, 6000072, 6000073, 6000074, 6000076, 6000077, 6000078, 6000079, 6000080, 6000081, 6000082, 6000083, 6000084, 6000085, 6000086, 6000087, 6000088, 6000089, 6000090, 6000091, 6000092, 6000093, 6000094, 6000095, 6000096, 6000097, 6000098, 6000099, 6000100, 6000101, 6000102, 6000103, 6000104, 6000105, 6000106, 6000107, 6000108, 6000109, 6000110, 6000111, 6000112, 6000113, 6000114, 6000115, 6000116, 6000117, 6000118, 6000119, 6000120, 6000121, 6000122, 6000123, 6000125, 6000126, 6000127, 6000128, 6000129, 6000130, 6000131, 6000132, 6000133, 6000134, 6000136, 6000138, 6000139, 6000140, 6000141, 6000142, 6000143, 6000144, 6000145, 6000146, 6000147, 6000148, 6000149, 6000150, 6000151, 6000152, 6000153, 6000154, 6000155, 6000156, 6000157, 6000158, 6000159, 6000160, 6000161, 6000162, 6000163, 6000164, 6000165, 6000166, 6000167, 6000168, 6000169, 6000170, 6000171, 6000172, 6000173, 6000174, 6000175, 6000176, 6000177, 6000178, 6000179, 6000180, 6000181, 6000182, 6000183, 6000184, 6000185, 6000186, 6000187, 6000188, 6000189, 6000190, 6000191, 6000192, 6000193, 6000194, 6000195, 6000196, 6000197, 6000198, 6000199, 6000200, 6000201, 6000202, 6000203, 6000204, 6000205, 6000206, 6000208, 6000209, 6000210, 6000211, 6000212, 6000213, 6000214, 6000215, 6000216, 6000217, 6000218, 6000219, 6000220, 6000221, 6000222, 6000223, 6000224, 6000225, 6000226, 6000227, 6000228, 6000229, 6000230, 6000231, 6000232, 6000233, 6000234, 6000235, 6000236, 6000237, 6000238, 6000239, 6000240, 6000241, 6000242, 6000243, 6000244, 6000245, 6000246, 6000247, 6000249, 6000250, 6000251, 6000252, 6000253, 6000254, 6000255, 6000256, 6000257, 6000258, 6000259, 6000260, 6000262, 6000263, 6000264, 6000265, 6000266, 6000267, 6000268, 6000269, 6000270, 6000271,6 000272, 6000273, 6000274, 6000275, 6000276, 6000277, 6000278, 6000279, 6000280, 6000281, 6000282, 6000283, 6000284, 6000285, 6000286, 6000287, 6000288, 6000290, 6000291, 6000292, 6000293, 6000294, 6000295, 6000296, 6000297, 6000298, 6000299, 6000300, 6000302, 6000303, 6000305, 6000306, 6000307, 6000308, 6000309, 6000310, 6000311, 6000312, 6000313, 6000314, 6000315, 6000316, 6000317, 6000318, 6000319, 6000320, 6000321, 6000322, 6000323, 6000324, 6000325, 6000326, 6000327, 6000328, 6000329, 6000330, 6000331, 6000332, 6000333, 6000334, 6000335, 6000336, 6000337, 6000338, 6000339, 6000341, 6000342, 6000343, 6000344, 6000345, 6000348, 6000349, 6000350, 6000351, 6000352, 6000353, 6000354, 6000355, 6000356, 6000357, 6000358, 6000359, 6000360, 6000361, 6000362, 6000364, 6000365, 6000366, 6000367, 6000368, 6000369, 6000370, 6000371, 6000372, 6000373, 6000375, 6000376, 6000377, 6000378, 6000379, 6000380, 6000381, 6000382, 6000383, 6000384, 6000385, 6000386, 6000387, 6000388, 6000389, 6000390, 6000392, 6340008, 11000021, 11000082, and 11000111.

Distribution pattern

Worldwide Distribution - USA including AZ, CA, CO, DC, DE, FL, GA, HI, IA, IL, KY, MA, MD, MN, MO, MS, MT, NC, NJ, ND, NE, NV, NY, OH, OK, OR, PA, SD, TX, VT, WA, and WI. Internationally to Canada, Algeria, Argentina, Australia, Austria, Belgium, Brazil, China, Czech Republic, Denmark, Finland, France, Gabon, Germany, India, Indonesia, Ireland, Italy, Japan, Korea Republic, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.