Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66603

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 20, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Midmark Corp dba Progeny Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The Vantage Digital Panoramic System delivers high-value features standard on every system: - Five pre-programmed exam settings, including bitewing, minimize exam set-up time and allow routine diagnostic exams to be performed extraorally for improved clinical efficiency and patient experience. - Progeny Clarity Enhanced Filter Set provides superior diagnostic imaging quality - Exclusive Focal Trough Alignment reminder assists in consistent patient positioning - Progeny Imaging software included for download on an unlimited number of operatory workstations with no additional costs - Exclusive VantageTrust remote support and training service offered at no charge for the first 60 days of use - Ceph upgradeable for expanded extraoral exam capabilities - The Progeny Vantage Panoramic System s proudly designed and manufactured in the USA; 2013 Midmark Corporation; 675 Heathrow Drive Lincolnshire, IL 60069; 847-415-9800 Toll-free 888-924-3800

Z-0651-2014
Recall number
Z-0651-2014
Initiated
September 20, 2013
Classification
Class II
Status
Terminated
Quantity
58 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
error was identified in the Real Time Controller (RTC) firmware

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Vantage Panoramic X-Ray systems. The error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. failure to stop").

Code information

Vantage Panoramic X-Ray System firmware versions: 3.0, 3.1, 3.2, and 3.3. Vantage Panoramic X-Ray System Catalog Numbers and Serial Numbers: Catalog Number V5000, Serial Numbers ZM81025, ZM81026, ZM81030, ZZ08005, ZZ08006, ZZ08007, ZZ08008, ZZ08009, ZZ08010, ZZ08011, ZZ08012, ZZ08013, ZZ08015, ZZ08016, ZZ08017, Z08018, ZZ08019, ZZ08020, ZZ08021, ZZ08022, ZZ08023, ZB51001, ZB51002, ZB51006, ZB51007, ZT18018, ZY49014, ZK58019, ZS54011, ZP01043, ZV78016, ZE09006, ZN44018, ZS59007, ZS59001, ZW15007, ZW34003, ZW34012, ZW34018, ZY49012, ZY49015, ZZ02010, ZM81022, ZS54012, ZE09019, ZM81017, ZT18004, ZT18015, ZN44020, ZE09012, ZM81023, ZT27012; Catalog Number V5050, Serial Number ZZ08014; Catalog Number V5100, Serial Numbers ZM81027, ZM81028, ZZ08003; Catalog Number V5150, Serial Number ZZ08002; Catalog Number V5150C, Serial Number ZM81016

Distribution pattern