Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 66698

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 18, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
SpineFrontier, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

PedFuse Reset Screw; Model: 01-80115-55 The PedFuse Reset pedicle screw implants intended to be used to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.

Z-0544-2014
Recall number
Z-0544-2014
Initiated
October 18, 2013
Classification
Class II
Status
Terminated
Recalling firm
SpineFrontier, Inc.
Quantity
35 Devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
During routine servicing, it was discovered that one 8mmx55mm Reset screw was assembled with an 8mm tulip head and a 9mm screw. The implant should have consisted of an 8mm tulip head and an 8mm screw. A recall was initiated to remove all remaining screws of the same lot from the field to investigate the nonconformity.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design (manufacturing process)

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

During routine servicing, it was discovered that one 8mmx55mm Reset screw was assembled with an 8mm tulip head and a 9mm screw. The implant should have consisted of an 8mm tulip head and an 8mm screw. A recall was initiated to remove all remaining screws of the same lot from the field to investigate the nonconformity.

Code information

Lot/Serial Numbers: 357305-000

Distribution pattern

US Distribution to VA and TX.