Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc., Princeton, NJ 08540 The i-Stat PT a prothrombin time test is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or Warfarin. The cartridge is to be used with the i-Stat 1 Analyzer and as part of the i-Stat System, the PT/INR test is to be used by trained health care professionals in accordance with a facility's policies and procedures.
- Recall number
- Z-0451-2014
- Initiated
- October 28, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Abbott Point Of Care Inc.
- Quantity
- 794,928 units US; 119,472 units OUS
App-derived interpretation
Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR.
Code information
K020355 List number 04J50-01, 04J50-02; 03P89-24 Lot number R13149A; R13151; R13151A and cartridge lots from C13139 up to and including C13270A
Distribution pattern
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Brazil, Chile, China, Finland, France, Germany, Greece, Hong Kong, Ireland, Italy, South Korea, Kuwait, Mexico, Netherland, New Zealand, Nigeria, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, south Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE, and United Kingdom,