Exact FDA event identifier

Event 66788

9 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 19, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Atrium Medical Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

9 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 9

Atrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with stopcock and in-line connectors; Product Code: 2050-000. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0735-2014
Recall number
Z-0735-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
10,448 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346. (All units in between and including these two numbers)

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.

device · product 2 of 9

Atrium Ocean Water Seal Chest Drain ATS Blood Recovery System with stopcock and in-line connectors Sterile fluid path packaging Product Code: 2050-070. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0736-2014
Recall number
Z-0736-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
624 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346. (All units in between and including these two numbers). To evacuate air and/or fluid from the chest cavity or mediastinum.

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.

device · product 3 of 9

Atrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with stopcock Blood Recovery System; Product Code: 2050-300. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0737-2014
Recall number
Z-0737-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
6,684 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346. (All units in between and including these two numbers)

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.

device · product 4 of 9

Atrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System Dual patient lines with stopcock and in-line connectors Product Code: 2052-000. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0738-2014
Recall number
Z-0738-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
361 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346. (All units in between and including these two numbers)

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.

device · product 5 of 9

Atrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System Dual patient lines with stopcock; .Product Code: 2052-300. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0739-2014
Recall number
Z-0739-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
1,002 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346. ( All units in between and including these two numbers)

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.

device · product 6 of 9

Atrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest Drain with in-line connectors Product Code: 3650-100. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0740-2014
Recall number
Z-0740-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
42,699 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346. (All units in between and including these two numbers)

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.

device · product 7 of 9

Atrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest Drain Dual patient lines; Product Code: 3652-100. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0741-2014
Recall number
Z-0741-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
954 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346. All units in between and including these two numbers.

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.

device · product 8 of 9

Atrium Express Dry Seal ATS (Autotransfusion) Blood Recovery System Chest Drain; Product Code: 4050-100N. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0742-2014
Recall number
Z-0742-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
7,274 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346 (All units in between and including these two numbers)

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.

device · product 9 of 9

Atrium Express Dry Seal ATS (Autotransfusion) Blood Recovery System Chest Drain with Prefilled Air Leak Monitor; Product Code: 4050-100P. To evacuate air and/or fluid from the chest cavity or mediastinum.

Z-0743-2014
Recall number
Z-0743-2014
Initiated
November 19, 2013
Classification
Class II
Status
Terminated
Quantity
1,944 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Code information

Lot number starting with 10883097 and the ending lot number 10913346 (All units in between and including these two numbers)

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.