device · product 1 of 1
AMSCO¿ Evolution¿ steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into and out of steam sterilizer units.
- Recall number
- Z-0478-2014
- Initiated
- October 18, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Steris Corporation
- Quantity
- 936 units
App-derived interpretation
The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.
Code information
Serial Number Range: 030061103 - 0535531220
Distribution pattern
Worldwide Distribution - USA Nationwide in the states of: AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MI, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and the District of Columbia including the countries of: Brazil, Canada, Chile, Guam, Mexico, Saudi Arabia, United Arab Emirates & Venezuela.