Event 67060

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 20, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Carestream Health Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Carestream DRX Evolution System (Standard Q) The DRX-Evolution System (Standard Q) is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstandbucky to complete x-ray exposures.

Z-0774-2014
Recall number
Z-0774-2014
Initiated
August 20, 2013
Classification
Class II
Status
Terminated
Recalling firm
Carestream Health Inc.
Quantity
4 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Carestream received a report of an exposure being initiated without a patient image captured. The patient was imaged 2 times before an image was captured. The field engineer who visited the site reported the system allowed an exposure without the assignment of a bucky or image receptor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Carestream received a report of an exposure being initiated without a patient image captured. The patient was imaged 2 times before an image was captured. The field engineer who visited the site reported the system allowed an exposure without the assignment of a bucky or image receptor.

Code information

Serial Numbers 5049, 5073, 5133, 5162

Distribution pattern

Nationwide Distribution including TX, IL, and GA.