Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67076

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 16, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
The Standing Company

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.

Z-0993-2014
Recall number
Z-0993-2014
Initiated
May 16, 2013
Classification
Class II
Status
Terminated
Recalling firm
The Standing Company
Quantity
approximately 119

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.

Code information

Serial No: 1306, 1313, 1342, 1349, 1362, 1358, 1359, 1392, 1391, 1406, 1402, 1407, 1426, 1441, 1451, 1018, 1453, 1399, 1523, 1477, 1480, 1482, 1497, 1494, 1505, 1516, 1533, 1537, 1536, 1535, 1542, 1544, 1549, 1543, 1550, 1562, 1563, 1566, 1580, 1572, 1573, 1574, 1571, 1578, 1583, 1641, 1593, 1591, 1592, 1599, 1604, 1600, 1615, 1616, 1614, 1622, 1631, 1637, 1643, 1635, 1640, 1634, 1648, 1666, 1652, 1657, 1663, 1664, 1676, 1689, 1665, 1683, 1691, 1688, 1701, 1703, 1711, 1698, 1712, 1719, 1728, 1731, 1726, 1740, 2165, 2172, 2175, 2177, 2257, 2182, 2197, 2205, 2258, 2259, 2262, 2270, 2283, 2279, 2281, 2298, 2306, 2310, 2312, 2330, 2334, 2503, 2515, 2328, 2514, 2523, 2525, 2550, 2561, 2568, 2581, 2573, 1684, and 1427.

Distribution pattern

Nationwide Distribution including AL, CA, DE, IA, IL , IN, KS, KY, MI, MO, NJ, OH, OK, TX, WA, WI,and WV.