Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67206

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 18, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Merit Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Radial Assist RAD BOARD, used to support the weight of a patient's arm and supplies for a medical procedure. The Merit RAD BOARD is a rigid and stiff body board intended for use for various medical purposes. RAD BOARD was specifically designed to support the weight of a patient's arm and supplies for medical procedure, in order to have optimal access to upper extremity vasculature, including radial and brachial arterial and venous access. The RAD BOARD is partially lined with a layer of lead-free Xenolite TB for additional radiation scatter protection.

Z-0901-2014
Recall number
Z-0901-2014
Initiated
December 18, 2013
Classification
Class II
Status
Terminated
Quantity
11

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Merit Medical Systems, Inc. is voluntarily recalling one lot (B507171) of RAD BOARD RB 100 devices. The affected devices are missing the main label which graphically depicts patient orientation on the board. The issue may allow the board to be oriented in the wrong direction where the semi-radiopaque material in the board (embedded Xenolite) could obscure patient anatomy during fluoroscopy. There

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Merit Medical Systems, Inc. is voluntarily recalling one lot (B507171) of RAD BOARD RB 100 devices. The affected devices are missing the main label which graphically depicts patient orientation on the board. The issue may allow the board to be oriented in the wrong direction where the semi-radiopaque material in the board (embedded Xenolite) could obscure patient anatomy during fluoroscopy. There

Code information

lot B507171

Distribution pattern

Worldwide Distribution - USA including NJ, TX, WV, MN, GA, and Internationally to Mexico and Japan.