device · product 1 of 1
RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Recall number
- Z-1081-2014
- Initiated
- November 25, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Elekta, Inc.
- Quantity
- 489
App-derived interpretation
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Code information
R4.2, R5.0, R5.1, R6.0, R6.1
Distribution pattern
Worldwide Distribution - USA including AL, CA, CT, GA, FL, IL, IN, KS, KY, LA, MD, MA, MI, MN, MO, NJ, NY, OH, PA, TN, TX, VA, and Internationally to Argentina, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cuba, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherland, Nigeria, Pakistan, Panama, Peru, Philippines, Portugal, Russia, Serbia, Solvenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Vietnam, and Yemen.