Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67406

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 05, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Integra LifeSciences Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Integra Cusa Excel Rx Only CEM" Nosecone for CUSA¿ Excel System Product Usage: Intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator The CUSA Electrosurgical Module (CEM) provides Desiccate Coagulation waveform electrosurgical capability to CUSA Excel handpieces. The CUSA handpiece with the CEM Nosecone works in conjunction with the Force FX generator in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.

Z-0994-2014
Recall number
Z-0994-2014
Initiated
December 05, 2013
Classification
Class II
Status
Terminated
Quantity
17253 boxes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position.

Code information

All unexpired lot numbers for CEM Nosecone Gatalouge No. C6623 and C6636 from 1/1/2010 to present. CEM" Nosecone, Catalogue No. C6623: 1083655, 1084219, 1084398, 1084399, 1084400, 1084682, 1084683, 1090937, 1091712, 1091714, 1091715, 1093125, 1093126, 1093128, 1093129, 1093471, 1093847, 1094575, 1094740, 1095587, 1095873, 1100237, 1100238, 1101217, 1101547, 1101548, 1102069, 1102070, 1102544, 1102545, 1102546, 1103102, 1103392, 1103444, 1103445, 1103446, 1104264, 1104265, 1105065, 1105066, 1105067, 1110208, 1110311, 1110348, 1110349, 1110473, 1110474, 1110892, 1111760, 1112182, 1112808, 1112809, 1113175, 1113176, 1113177, 1113178, 1113662, 1114112, 1114654, 1114655, 1114656, 1115224, 1115225, 1115499, 1115500, 1115501, 1115853, 1115854, 1120771, 1120772, 1120773, 1120774, 1121141, 1121142, 1121383, 1121627, 1121628, 1122257, 1122259, 1122260, 1123050, 1123312, 1123915, 1124642, 1124643, 1124980, 1125393, 1125394, 1125933, 1130036, 1130037, 1130038, 1130039, 1130218, 1130219, 1130663, 1130664, 1131340, 1131344, 1131986, 1132207, 1132326, 1132327, 1132328, 1132880 CEM" Nosecone, Catalogue No. C6636: 1084216, 1084293, 1084377, 1084680, 1084681, 1090123, 1090124, 1090125, 1090126, 1090660, 1090661, 1090826, 1091395, 1091434, 1091700, 1092388, 1092389, 1092390, 1092391, 1092392, 1093012, 1093013, 1093130, 1093131, 1093472, 1093473, 1093474, 1093824, 1093825, 1093826, 1093827, 1094741, 1095104, 1095105, 1095106, 1095414, 1095415, 1095418, 1095419, 1100239, 1100240, 1100241, 1101220, 1101221, 1102647, 1102648, 1102649, 1102650, 1102651, 1103103, 1103394, 1103395, 1103437, 1103848, 1103851, 1103916, 1103917, 1103919, 1103920, 1103921, 1103922, 1104258, 1104259, 1104260, 1104261, 1104262, 1105068, 1105069, 1105070, 1110333, 1110975, 1110976, 1110977, 1111491, 1111492, 1111493, 1112186, 1112187, 1112211, 1112212, 1112769, 1112770, 1112771, 1113180, 1113181, 1113182, 1113183, 1113184, 1113664, 1113665, 1113666, 1114358, 1114359, 1114657, 1114658, 1114659, 1114660, 1115036, 1115037, 1115038, 1115039, 1115503, 1115504, 1115505, 1115855, 1115856, 1115857, 1115858, 1115859, 1115860, 1120360, 1120361, 1120362, 1120775, 1120776, 1120777, 1121568, 1121729, 1122344, 1122346, 1122347, 1123053, 1123054, 1123497, 1123498, 1123499, 1123500, 1123501, 1124054, 1124055, 1124703, 1124704, 1124705, 1124706, 1124707, 1124710, 1125395, 1125396, 1125397, 1125398, 1125399, 1125400, 1125476, 1125934, 1125935, 1125936, 1125937, 1125938, 1125939, 1130220, 1130221, 1130222, 1130223, 1130834, 1131243, 1131987, 1132777, 1132778, 1132779, 1132780, 1132879, 1133061, 1103850, 1103852, 1103917, 1103918, 1103922, 1104257, 1104263, 1105071, 1110475, 1110476, 1110978, 1111490, 1111493, 1112187, 1112188, 1112212, 1113179, 1113180, 1113663, 1114357, 1114358, 1114659, 1114660, 1115039, 1115502, 1115504, 1115505, 1120776, 1120777, 1121167, 1121168, 1121169, 1121565, 1121566, 1121567, 1122344, 1122743, 1123051, 1123052, 1123497, 1123498, 1123499, 1123500, 1123501, 1123508, 1123509, 1123510, 1123511, 1123512, 1124052, 1124053, 1124707, 1124709

Distribution pattern

Worldwide Distribution - USA Nationwide