Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67445

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 27, 2014
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Cardiovascular Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

CSI, Diamondback 360 Coronary Orbital Atherectomy System, Part Number 70058-02, Model Number DBEC-125 Product Usage: The system is indicated to facilitate stent delivery in patients with coronary artery disease (CAD) who are acceptable candidates for PTCA or stenting due to de novo, severely calcified coronary artery lesions.

Z-1095-2014
Recall number
Z-1095-2014
Initiated
January 27, 2014
Classification
Class III
Status
Terminated
Quantity
318

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
does not meet specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Cardio vascular Systems Inc. has initiated a recall to request the immediate removal and return of Diamondback 360 Coronary Orbital Atherectomy Device (OAD). The device mistakenly contains a saline line that is used in our peripheral atherectomy device. This saline line was not part of the FDA approval for use in the coronary device. It therefore does not meet specification for this device. C

Code information

Lot: 85408 -85412, 85585 -85587, 87276, 87278, 87280, 87317, 87462, 87464, 87793, 87949 - 87952, 88520 - 88524, 88975 - 88979, 89498, 89501- 89504, 89952, 89954, 89956, 91876, 91877, 91891, 92124, 92156.

Distribution pattern