Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67505

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 12, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Biosense Webster, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

CartoUnivu Module, Model: Carto 3 Software versions 3.2.2 and 3.2.3; Catalog No. KT-5400-124 WITH THE CARTOUNIVU MODULE, FLUOROSCOPIC CAPTURES (IMAGES AND CINE CLIPS) CAN BE IMPORTED TO AND DISPLAYED IN THE CART0 3 SYSTEM. THESE IMAGES ARE NOT FOR DIAGNOSTIC PURPOSES. WHEN FLUOROSCOPIC CAPTURES ARE INTEGRATED INTO THE CART0 3 SYSTEM MAP VIEWERS (MAP VIEWER AND ADDITIONAL VIEW), CART0 3 MAPS, MRl OR CT IMAGES, AND CATHETER VISUALIZATION ARE SUPERIMPOSED ON THE FLUOROSCOPIC CAPTURE.

Z-1045-2014
Recall number
Z-1045-2014
Initiated
February 12, 2014
Classification
Class II
Status
Terminated
Recalling firm
Biosense Webster, Inc.
Quantity
8 units total (5 units in US)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Biosense Webster has initiated a recall of the CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3) when used with Siemens Axiom Artis VB Fluoroscopy systems in particular procedural configurations only. If the table rotation is not returned to zero point, there is a misalignment between the Carto 3 System map display and the fluoroscopic capture.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Biosense Webster has initiated a recall of the CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3) when used with Siemens Axiom Artis VB Fluoroscopy systems in particular procedural configurations only. If the table rotation is not returned to zero point, there is a misalignment between the Carto 3 System map display and the fluoroscopic capture.

Code information

Serial No. 13201, 11125, 11519, 11320, 11115.

Distribution pattern

Worldwide Distribution - USA including CA, KY, NJ, MA, and UT and Internationally to Belgium, Austria, and Germany.