Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67517

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 26, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Mar Cor Purification

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..

Z-1213-2014
Recall number
Z-1213-2014
Initiated
June 26, 2013
Classification
Class II
Status
Terminated
Recalling firm
Mar Cor Purification
Quantity
232 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.

Code information

Serial Numbers: 1294627, 1298383, 1298386, 1299352, 1299353, 1299354, 1299355, 1299356, 1299486, 1299487, 1299488, 1299489, 1299503, 1299572, 1299712, 1299713, 1299716, 1299718, 1299719, 1299720, 1299740, 1299742, 1299777, 1299779, 1299780, 1299831, 1299835, 1299850, 1299867, 1299868, 1299895, 1299897, 1299992, 1299993, 1299994, 1299996, 1299997, 1300138, 1300224, 1300248, 1300363, 1300364, 1300367, 1300368, 1300558, 1300560, 1300561, 1300562, 1300564, 1300565, 1300621, 1300649, 1300650, 1300700, 1300701, 1301081, 1301116, 1301360, 1301361, 1301386, 1301888, 1301990, 1301992, 1301997, 1302030, 1302032, 1302149, 1302214, 1302226, 1302342, 1302345, 1302346, 1302385, 1302430, 1302447, 1302448, 1302452, 1302456, 1302464, 1302517, 1302555, 1302558, 1302642, 1302696, 1302706, 1302708, 1302748, 1302767, 1302768, 1302810, 1302813, 1302922, 1302923, 1302924, 1302925, 1302926, 1302927, 1302987, 1303005, 1303108, 1303315, 1303316, 1303346, 1303348, 1303351, 1303352, 1303505, 1303516, 1303532, 1303533, 1303534, 1303557, 1303558, 1303625, 1303626, 1303689, 1303690, 1303693, 1303702, 1303711, 1303768, 1303781, 1303799, 1303806, 1303821, 1303872, 1303877, 1303908, 1303920, 1303975, 1303976, 1304070, 1304076, 1304077, 1304081, 1304163, 1304166, 1304167, 1304170, 1304173, 1304174, 1304194, 1304196, 1304203, 1304273, 1304293, 1304294, 1304297, 1304333, 1304340, 1304343, 1304344, 1304346, 1304349, 1304376, 1304381, 1304501, 1304502, 1304503, 1304507, 1304508, 1304551, 1304585, 1304587, 1304724, 1304725, 1304775, 1304778, 1304779, 1304829, 1304870, 1304902, 1304903, 1304929, 1305006, 1305007, 1305038, 1305040, 1305041, 1305087, 1305088, 1305089, 1305131, 1305133, 1305210, 1305238, 1305239, 1305355, 1305363, 1305608, 1305609, 1305650, 1305651, 1305655, 1305698, 1305700, 1305701, 1305717, 1305732, 1305735, 1305750, 1305751, 1305752, 1305795, 1305928, 1305931, 1305932, 1305975, 1305995, 1306007, 1306042, 1306046, 1306078, 1306125, 1306154, 1306178, 1306209, 1306215, 1306226, 1306259, 1306322, 1306341, 1306428, 1306429, 1306430, 1306431, 1306432, 1306654, 1306662, 1306663, 1306664, 1306665.

Distribution pattern

USA Nationwide Distribution including Puerto Rico.