Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67775

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 18, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Instruments Div. of Stryker Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

2.5MM Aggressive Blade Known as a Sinuscopic Shaver Blade, this device is composed of a metal cutter and plastic drive shaft that goes into a metal housing supported by a plastic hub. ENT Precision Cutter, Aggressive, ESSxlHUMMER 4, Rx Only It is used for ENT procedures and functions by means of rotation when adapted to a handpiece and console. Through the rotation, the necessary tissue is resected as required by the user. The blade housing has an etching line guide that starts with a straight line and along with a scale from 1cm to 7cm in 1cm increments that serves the doctor as a guide by indicating how far the device is being introduced into the surgical area.

Z-1247-2014
Recall number
Z-1247-2014
Initiated
March 18, 2014
Classification
Class II
Status
Terminated
Quantity
90 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Stryker Instruments is voluntarily recalling the single use 2.5MM Aggressive Blade because the line on the affected part (P/N 5290-928-000, Lot #13280CG2) was a straight line without a number scale or logo. The line required for this part number includes a number scale that goes from 1cm  7cm.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Stryker Instruments is voluntarily recalling the single use 2.5MM Aggressive Blade because the line on the affected part (P/N 5290-928-000, Lot #13280CG2) was a straight line without a number scale or logo. The line required for this part number includes a number scale that goes from 1cm  7cm.

Code information

P/N 5290-628-000, Lot #13280CG2

Distribution pattern

US Distribution including the states of NV, CA, WA, TN, OR, SD, ID, IN and CO., and Internationally to Malaysia