Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 67837

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 27, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems (Cleveland) Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core 128 and Ingenuity Flex scanners are Computed Tomography X-ray systems.

Z-1461-2014
Recall number
Z-1461-2014
Initiated
March 27, 2014
Classification
Class II
Status
Terminated
Quantity
135

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.

Code information

Brilliance CT 16: Model Number: 728246; Serial Number: 5314, 6719, 6726, 6727, 6728, 6729, 6730, 6731, 6732, 6734, 6735, 6736, 6737, 6738, 6744, 6746, 6747, 6748, 6749, 6750, 6751, 6752, 6753, 6754, 6756, 6757, 6758, 6759, 6760, 6761, 6762, 6764, 6767, 6768, 6772, and 50213. Brilliance CT 64: Model Number: 728231; Serial Number: 10805, 10806, 90047, 95880, 95881, 95882, 95885, 95892, 95903, 95907, 95910, 95912, 95913, 95914, 95915, 95918, 95920, 95921, 95922, and 98895. Brilliance CT Big Bore Oncology: Model Number: 728243; Serial Number: 7971, 7977, 7980, 7989, and 7994. Brilliance CT Big Bore Radiology: Model Number: 728244; Serial Number: 7957, 7965, 7979, and 7985. Brilliance iCT: Model Number: 728306; Serial Number: 100477, 100483, 100487, 100489, 100491, 100492, 100494, 100495, 100496, 100500, 100502, 100504, and 100507. Ingenuity CT: Model Number: 728326; Serial Number: 300165 and 333013. Ingenuity Core: Model Number: 728321; Serial Number: 310138, 310142, 310144, 310145, 310149, 310151, 310152, 310153, 310154, 310155, 310158, 310160, 310161, 310162, 310163, 310164, 333013, 333015, and 333017. Ingenuity Core128: Model Number: 728323; Serial Number: 320113, 320114, 320116, 320117, 320118, 320120, 320121, 320122, 320124, 320125, 320128,320129, 320130, 320134, 320137, 320138, 320139, 320140, 320141, 320142, 320143, 320144,320145, 320146, 320147, 320148, 336009, 336010, 336011, 336012, 336013, 336015, 336016, 336018, and 336023. Ingenuity Flex: Model Number: 728317; Serial Number: 345005

Distribution pattern

Nationwide Distribution - USA including the states of CA, FL, IL, IN, KY, MD,SC, and TX.