April 02, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Philips HeartStart XL+ Defibrillator/Monitor with system software version B.00.00 or B.00.01 installed Model: 861290, automatic external defibrillator.

Z-1438-2014
Recall number
Z-1438-2014
Initiated
April 02, 2014
Classification
Class II
Status
Terminated
Quantity
9872 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software communication failure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software communication failure may occur on the HeartStart XL+ locking the user out of clinical mode and possibly causing a delay in therapy.

Code information

USO1100106 to USO1100372 USN1100376 to USN1100960 USD1100961 to USD1101095 US11201096 to US11201186 US21201187 to US21201239 US31201240 to US31201537 US41201538 to US41201585 US51201586 to US51201721 US61201722 to US61201924 US71201925 to US71202048 US81202049 to US81202168 US91202169 to US91202514 USO1202515 to USO1202990 USN1202991 to USN1203537 USD1203538 to USD1203968 US11303969 to US11303972 US21303980 to US21304488 US31304489 to US31305042 US41305043 to US41305295 US51305296 to US51305450 US61305451 to US61306138 US71306139 to US71306542 US81306543 to US81306998 US91306999 to US91307261 USO1307306 to USO1307458 USN1307616 to USN1308423 USD1308424 to USD1309471 US11409472 to US11410212 US21410213 to US21410563

Distribution pattern

Worldwide Distribution - USA (nationwide) including the countries of: ARGENTINA AUSTRALIA AUSTRIA BELGIUM CANADA CHINA CZECH REPUBLIC DENMARK FRANCE GERMANY ITALY LATVIA NETHERLANDS NEW ZEALAND NORWAY PHILIPPINES PORTUGAL SOUTH KOREA SPAIN SWITZERLAND THAILAND TURKEY and UNITED KINGDOM.