device · product 1 of 1
Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), consisting of the Monitor and test strip, is used for quantitative determination of international normalized ratio (INR) in fresh, capillary whole blood to monitor the effect of warfarin therapy on clotting time by health care professionals. (in vitro diagnostic use).
- Recall number
- Z-1546-2014
- Initiated
- April 16, 2014
- Classification
- Class I
- Status
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Quantity
- Estimated 99,795 Total units
App-derived interpretation
Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Code information
Lot/Unit Codes: Applicable to all Alere INRatio2 PT/INR Professional Test Strips lots packaged from 8/22/13 through 3/28/14, including the following: 324810, 324811, 324812, 327543, 327544, 327549, 328109, 328705, 328707, 330238, 330852, 331340, 331916, 331922, 332841, 332842, 332843, 332959, 333362, 333654, 333655, 333864, 333865, 334572, 334575, 335552, 335553, 335796, 335797, 336426, 336668, 337350, 338965, 338966, 339257, 339707, 327856V Expiration date, or Expected shelf life: range from 4/30/14 to 1/31/15
Distribution pattern
US Nationwide Distribution