Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68047

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 03, 2014
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Hospira Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.

Z-0095-2015
Recall number
Z-0095-2015
Initiated
February 03, 2014
Classification
Class I
Status
Terminated
Recalling firm
Hospira Inc.
Quantity
978 power supplies

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.

Code information

Wall Mount GemStar Power Supply List Number: 13026; Lot Numbers: 25283G1, 26286G1, 26287G1, 27292G1, 27294G1, 27295G1, 27296G1

Distribution pattern

Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and, the countries of Austria, Belgium, Brazil, Columbia, France, Germany, Greece, Netherlands, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

device · product 2 of 2

The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.

Z-0096-2015
Recall number
Z-0096-2015
Initiated
February 03, 2014
Classification
Class I
Status
Terminated
Recalling firm
Hospira Inc.
Quantity
US Distribution: 4,709 power supplies; Foreign Distribution: 7,425 power supplies

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.

Code information

Desktop GemStar Power Supply List Number: 13072-05; 1) US Distribution Lot Numbers: 10115G1, 11121G1, 11122G1, 1123G1, 11124G1, 12143G1, 12144G1, 12178G1, 15185G1, 12195G1, 17214G1, 18217G1, 18219G1, 19225G1, 20233G1, 21242G1, 22251G1, 23259G1, 24271G1, 25280G1, 26289G1, 26290G1, 27293G1, 27297G1, 27300G1; 2) Foreign Distribution Lot Numbers: 11128G1, 12140G1, 12142G1, 12148G1, 14176G1, 15187G1, 15196G1, 16207G1, 16208G1, 18218G1, 19226G1, 19229G1, 20234G1, 21241G1, 22255G1, 23262G1, 24270G1, 25276G1, 26285G1, 27299G1, 27301G1

Distribution pattern

Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and, the countries of Austria, Belgium, Brazil, Columbia, France, Germany, Greece, Netherlands, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom.