device · product 1 of 1
Recalled product is a component of the Newport Medical Instruments e360 Ventilator: single board computer, part number SBC2105A. Intended to provide invasive or noninvasive ventilator support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency.
- Recall number
- Z-1577-2014
- Initiated
- April 18, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Nellcor Puritan Bennett Inc. (dba Covidien LP)
- Quantity
- 5,765
App-derived interpretation
Reports of the audible alarm not sounding when the ventilator is in an alarm state. The ventilator, as well as the visual and remote alarms, continue to operate normally.
Official device-enrichment evidence · Sourced
Component design/selection
Inspect official wording and provenance
Reason for recall
Reports of the audible alarm not sounding when the ventilator is in an alarm state. The ventilator, as well as the visual and remote alarms, continue to operate normally.
Code information
Part Number: SBC2105A. Model Number(s): e360 Ventilator. Serial numbers N11360719000 through N14360426029.
Distribution pattern
Worldwide Distribution -- USA, including the states of Arizona, California, Colorado, Connecticut, Florida, Georgia, Kentucky, Massachusetts, Maryland, Missouri, Nevada, Pennsylvania, Rhode Island, Texas, Virginia, and Washington; and, the countries of Algeria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Dominican Republic, Ecuador, Egypt, England, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Japan, Kazakhstan, Kenya, Lebanon, Liechtenstein, Lithuania, Mexico, Morocco, Namibia, Nepal, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Republic of Armenia, Republic of KOS, Russia, Saudi Arabia, Singapore, Slovenia, S. Africa, Sri Lanka, Taiwan, Thailand, Tunisia, Turkey, UAE, Uruguay, Vietnam, and Yemen.