Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68109

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 08, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
The Anspach Effort, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ANSPACH Cutting bur, 3 mm Diamond Ball, Standard Length. REF L-3SD Use with LONG Attachments. Used for cutting and shaping bone, including the spine and cranium.

Z-1574-2014
Recall number
Z-1574-2014
Initiated
January 08, 2014
Classification
Class II
Status
Terminated
Recalling firm
The Anspach Effort, Inc.
Quantity
30 cutting burs affected

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Anspach product code L-3SD contained product code S-3SD which is shorter in length than the labeled product.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling mix-ups

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Anspach product code L-3SD contained product code S-3SD which is shorter in length than the labeled product.

Code information

Lot/Set/Serial: G323081803.

Distribution pattern

Nationwide Distribution including FL, TX, NC, IL, CO, and PA.