Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68259

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 21, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
LAP of America Laser Applications, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.

Z-1866-2014
Recall number
Z-1866-2014
Initiated
March 21, 2014
Classification
Class II
Status
Terminated
Quantity
1 system

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.

Code information

Product Code: 0007569-0001

Distribution pattern

North Carolina