Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68286

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 12, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ortho-Clinical Diagnostics

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701 -- For in vitro diagnostic use only --- VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent.

Z-1804-2014
Recall number
Z-1804-2014
Initiated
May 12, 2014
Classification
Class II
Status
Terminated
Quantity
Domestic - 612 units, Foreign - 461 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.

Code information

Lots 1732 (exp. 31-Jul-2014), 1743 (exp. 31-Jul-2014), 1753 (exp. 28-Feb-2015), 1764 (31-Jul-2015)

Distribution pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.

device · product 2 of 2

VITROS (R) Chemistry Products FS Calibrator 1, Product Code/ Catalog Number 6801873 -- For in vitro diagnostic use only --- VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate the VITROS 5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Z-1805-2014
Recall number
Z-1805-2014
Initiated
May 12, 2014
Classification
Class II
Status
Terminated
Quantity
Domestic - 464 units; Foreign - 361 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.

Code information

Lots 91131 (exp. 07-Mar-2015) and 91241 (exp. 28-Feb-2016)

Distribution pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.