device · product 1 of 2
Philips MR systems: Ingenia HA FLEX TRAK Trolley (for Ingenia 1.5T & 3.0T products) Product Usage: Diagnostic MR System. The Height Adjustable FlexTrak trolley and Trolley Variable Height IRF is an optional device designed to prepare the patient outside the examination room to enhance workflow and make additional non-magnetic hospital beds unnecessary.
- Recall number
- Z-1826-2014
- Initiated
- May 15, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Philips Medical Systems, Inc.
- Quantity
- 420
App-derived interpretation
Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF.
Code information
Model Numbers: 9896 030 1929.2, 9896 030 1929.3
Distribution pattern
Worldwide Distribution - US (nationwide), Canada, Mexico Australia, Austria, Germany, Belarus, Bulgaria, Brazil, China, Denmark, Estonia, Finland, France, Germany, Greece, Italy, Jordan, Korea, Republic of Korea, Netherlands, New Zealand, Norway, Oman, Pakistan, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.