device · product 1 of 1
Stirling Ultracold Portable Shuttle, ULT-25. Storage for cold-chain management, short-term or long-term ultra-low and storage and transport of biological specimens.
- Recall number
- Z-1940-2014
- Initiated
- March 20, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Global Cooling Inc
- Quantity
- 413 units
App-derived interpretation
The external power supply input is overheating at the connection to the freezer.
Official device-enrichment evidence · Sourced
Use error
Inspect official wording and provenance
Reason for recall
The external power supply input is overheating at the connection to the freezer.
Code information
Codes: ULT-25-255 & ULT-25-713
Distribution pattern
Worldwide Distribution -- USA, including the states of AK, AR, AZ, CA, CO, CT, FL, GA, GU, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, and WA; and, the countries of Australia, Brazil, Canada, China, India, Indonesia, Japan, Korea, Kuwait, Mexico, Netherlands, Pakistan, Singapore, Switzerland, Taiwan, Thailand, United Arab Emirates, and United Kingdom.