device · product 1 of 1
RUSCH, LaserTube (Rubber), Laser Resistant Tracheal Tube, cuffed, Teleflex Medical, Willy-Rusch, GmbH. A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
- Recall number
- Z-1851-2014
- Initiated
- May 23, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Teleflex Medical
- Quantity
- 1658 ea
App-derived interpretation
Although the Natural Rubber Latex caution symbol is clearly displayed on the product label, the accompanying caution statement "Caution: This product contains Natural Rubber Latex which may cause Allergic Reactions" was not printed on the product label.
Official device-enrichment evidence · Sourced
Error in labeling
Inspect official wording and provenance
Reason for recall
Although the Natural Rubber Latex caution symbol is clearly displayed on the product label, the accompanying caution statement "Caution: This product contains Natural Rubber Latex which may cause Allergic Reactions" was not printed on the product label.
Code information
Product Code: 102004040, Lot #13371, 13381, 13291, 13281, 13251 & 13231; Product Code: 102004050, Lot #13461, 13441, 13401, 13431, 13421, 13381, 13371, 1329, 13301, 13311, 13361, 13261, 13231, 13251 & 13241; Product Code: 102004060, Lot #13441, 13451, 13431, 13411, 13381, 13361, 13311, 13301, 13281, 13271, 13261, 13241 & 13231; Product Code: 102004070, Lot #13441, 13421, 13381, 13351, 13291, 13311, 13271 & 13261; Product Code: 102004070, Lot #13251 & 13241; and Product Code: 102004080, Lot #13371, 13301 & 13251.
Distribution pattern
Nationwide Distribution and the country of Colombia.