Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68489

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 27, 2014
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Vascular Solutions, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Langston Dual Lumen Catheters, 6Fr, REF 5550 and 5540. The Vascular Solutions Langston Dual Lumen Pressure Monitoring catheter is used as a pressure measurement catheter and for delivery of contrast media during angiographic studies.

Z-1981-2014
Recall number
Z-1981-2014
Initiated
May 27, 2014
Classification
Class I
Status
Terminated
Recalling firm
Vascular Solutions, Inc.
Quantity
8,580 total (US 5714)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The inner catheter of some Langston V2 Dual Lumen Catheters have separated from the device hub during contrast pressure injections. This may require an intervention to retrieve the separated piece from the patient to prevent injury.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The inner catheter of some Langston V2 Dual Lumen Catheters have separated from the device hub during contrast pressure injections. This may require an intervention to retrieve the separated piece from the patient to prevent injury.

Code information

Lot Numbers: 569436, 569437, 569962, , 569963, 569964, 569965, 570177, 570178, 570337, 570338, 570339, 570340, 570341, 570504, 570505, 570506, 570507, 570592, 570593, 570674, 570675, 570744, 570745, 570746, 570836, 570926, 570927, 570928, 570929, 571099, 571100, 571101, 571102, 571103, 571350, 571412, 571413, 571414, 571415, and 571727.

Distribution pattern

Worldwide Distribution - USA nationwide including DC and Puerto Rico, AUSTRIA, CANADA, GERMANY, ISRAEL, ITALY, NETHERLANDS, SPAIN, SWEDEN SWITZERLAND, and UNITED KINGDOM.