device · product 1 of 1
Langston Dual Lumen Catheters, 6Fr, REF 5550 and 5540. The Vascular Solutions Langston Dual Lumen Pressure Monitoring catheter is used as a pressure measurement catheter and for delivery of contrast media during angiographic studies.
- Recall number
- Z-1981-2014
- Initiated
- May 27, 2014
- Classification
- Class I
- Status
- Terminated
- Recalling firm
- Vascular Solutions, Inc.
- Quantity
- 8,580 total (US 5714)
App-derived interpretation
The inner catheter of some Langston V2 Dual Lumen Catheters have separated from the device hub during contrast pressure injections. This may require an intervention to retrieve the separated piece from the patient to prevent injury.
Official device-enrichment evidence · Sourced
Component design/selection
Inspect official wording and provenance
Reason for recall
The inner catheter of some Langston V2 Dual Lumen Catheters have separated from the device hub during contrast pressure injections. This may require an intervention to retrieve the separated piece from the patient to prevent injury.
Code information
Lot Numbers: 569436, 569437, 569962, , 569963, 569964, 569965, 570177, 570178, 570337, 570338, 570339, 570340, 570341, 570504, 570505, 570506, 570507, 570592, 570593, 570674, 570675, 570744, 570745, 570746, 570836, 570926, 570927, 570928, 570929, 571099, 571100, 571101, 571102, 571103, 571350, 571412, 571413, 571414, 571415, and 571727.
Distribution pattern
Worldwide Distribution - USA nationwide including DC and Puerto Rico, AUSTRIA, CANADA, GERMANY, ISRAEL, ITALY, NETHERLANDS, SPAIN, SWEDEN SWITZERLAND, and UNITED KINGDOM.