device · product 1 of 1
The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and Hospira infusion pumps. The MMS provides trained healthcare professionals with the capability to send, receive, report, and store information from interfaced external systems, and to configure and edit infusion programming parameters. The MMS is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by a trained healthcare professional according to physicians orders.
- Recall number
- Z-2027-2014
- Initiated
- June 06, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Hospira Inc.
- Quantity
- 13 units
App-derived interpretation
software defect
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" on an unmatched medication will produce a
Code information
1) Software Version 5.5, List Number: 16037-04-13; 2) Software Version 5.8.1, List Number: 16037-04-31; 3) Software Version 5.8.2, List Number: 16037-04-33
Distribution pattern
US Nationwide in the states of FL, LA, PA, WA, WY and the country of Hong Kong