Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68838

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 11, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medtronic Neuromodulation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medtronic Ascenda Intrathecal Catheter models 8780 and 8781, and Ascenda Intrathecal Catheter Pump Segment Revision Kit, model 8784. The Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Ascenda Intrathecal Catheter.

Z-2172-2014
Recall number
Z-2172-2014
Initiated
July 11, 2014
Classification
Class II
Status
Terminated
Recalling firm
Medtronic Neuromodulation
Quantity
1200 (551 US, 649 OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria. There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria. There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne

Code information

Serial Numbers are in the following format: "N" + 9 digits (e.g. N12345001. Product beginning with the following letter/number combinations are affected: N464184, N464200, N464206, N464254, N464257, N464258, N464262, N464263, N464264, N464271, N464693, N464703, N464708, N464725, N464727, N464730, N464734, N464736, N464738, N464742, N464746, N464750, N464752, N464841, N464924, N464930, N464995, N464995, N465020, N465038, N465063, N465065, N465073, N465082, N465088, N465091, N465094, N465111, N465125, N465129, N465131, N465136, N465138, N465144, N465145, N465146, N465148, N465150, N466065, N466067, N466067, N466068, N466069, N466071, N466072, N466075, N466134, N466167, N466177, N466196, N466201, N466240, N466244, N466259, N466289, N466290, N466291, N466306, N466368, N466390, N466392, N466394, N466398, N466399, N466631, N466639, N466645, N466678, N466678, N466754, N466758, N466761, N466886, N466897, N466901, N466902, N466911, N466914, N467103, N467109, N467113, N467114, N467116, N467126, N467138, N467171, N467180, N467183, N467192, N467193, N467194, N467195, N467197, N467198, N467203, N467204, N467206, N467209, N467351, N467351, N467352, N467355, N467356, N467357, N467395, N467399, N467401, N467402, N467407, N467444, N467451, N467453, N467454, N467563, N467568, N467570, N467601, N467606, N467608, N467609, N467611, N467612, N467614, N467618, N467621, N467626, N468288, N468343, N468401, N468404, N468406, N468406, N468448, N468494, N468500, N468507, N468510, N468511, N468514, N468516, N468519, N468522, N468574, N468580, N468621, N468622, N468624, N468626, N468641, N469182, N469262, N469332, N469360, N469360, N469388, N469420, N469422, N469428, N469431, N469433, N469434, N469436, N469438, N469440, N469446, N469447, N469449, N469452, N469460, N469465, N469466, N469469, N469474, N469478, N469507, N469511, N469530, N469533, N469549, N469579, N469583, N469586, N469588, N470163, N470197, N470234, N470303, N470307, N470401, N470408, N470414, N470420, N470424, N470453, N470459, N470919, N470923, N470929, N470931, N470931, N470934, N470938, N470940, N470941, N470951, N470953, N470954, N471442, N471443, N471446, N471447, N471448, N471449, N471487, N471488, N471490, N471490, N471491, N471764, N471791, N471840, N471853, N471856, N471860, N471863, N471864, N471865, N471866, N471870, N471871, N471875, N471880, N472310, N472326, N472333, N472336, N472342, N472345, N472347, N472857, N472861, N472866, N472868, N472873, N472875, N472876, N472878, N472882, N472885, N472886, N472890, N472894, N472900, N472905, N473140, N473144, N473148, N473154, N473156, N473176, N473178, N473180, N473202, N473207, N473211, N473221, N473222, N473223, N473228, N473230, N473311, N473311, N473323, N473377, N473401, N473426, N473438, N473492, N473493, N473989, N473996, N473999, N474001, N474005, N474006, N474007, N474010, N474018, N474019, N474023, N474027, N474037, N474038, N474047, N474047, N474421, N474423, N474427, N474431, N474433.

Distribution pattern

Worldwide Distribution - USA (nationwide) including the states of AL,AR,AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, Ml, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WV., and the countries of Austria, Belgium, Cyprus, Denmark, France, Germany, Iraq, Italy, Lebanon, Netherlands, Panama, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, United Kingdom and Uruguay.