device · product 1 of 1
Medtronic Ascenda Intrathecal Catheter models 8780 and 8781, and Ascenda Intrathecal Catheter Pump Segment Revision Kit, model 8784. The Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Ascenda Intrathecal Catheter.
- Recall number
- Z-2172-2014
- Initiated
- July 11, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 1200 (551 US, 649 OUS)
App-derived interpretation
Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria. There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria. There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne
Code information
Serial Numbers are in the following format: "N" + 9 digits (e.g. N12345001. Product beginning with the following letter/number combinations are affected: N464184, N464200, N464206, N464254, N464257, N464258, N464262, N464263, N464264, N464271, N464693, N464703, N464708, N464725, N464727, N464730, N464734, N464736, N464738, N464742, N464746, N464750, N464752, N464841, N464924, N464930, N464995, N464995, N465020, N465038, N465063, N465065, N465073, N465082, N465088, N465091, N465094, N465111, N465125, N465129, N465131, N465136, N465138, N465144, N465145, N465146, N465148, N465150, N466065, N466067, N466067, N466068, N466069, N466071, N466072, N466075, N466134, N466167, N466177, N466196, N466201, N466240, N466244, N466259, N466289, N466290, N466291, N466306, N466368, N466390, N466392, N466394, N466398, N466399, N466631, N466639, N466645, N466678, N466678, N466754, N466758, N466761, N466886, N466897, N466901, N466902, N466911, N466914, N467103, N467109, N467113, N467114, N467116, N467126, N467138, N467171, N467180, N467183, N467192, N467193, N467194, N467195, N467197, N467198, N467203, N467204, N467206, N467209, N467351, N467351, N467352, N467355, N467356, N467357, N467395, N467399, N467401, N467402, N467407, N467444, N467451, N467453, N467454, N467563, N467568, N467570, N467601, N467606, N467608, N467609, N467611, N467612, N467614, N467618, N467621, N467626, N468288, N468343, N468401, N468404, N468406, N468406, N468448, N468494, N468500, N468507, N468510, N468511, N468514, N468516, N468519, N468522, N468574, N468580, N468621, N468622, N468624, N468626, N468641, N469182, N469262, N469332, N469360, N469360, N469388, N469420, N469422, N469428, N469431, N469433, N469434, N469436, N469438, N469440, N469446, N469447, N469449, N469452, N469460, N469465, N469466, N469469, N469474, N469478, N469507, N469511, N469530, N469533, N469549, N469579, N469583, N469586, N469588, N470163, N470197, N470234, N470303, N470307, N470401, N470408, N470414, N470420, N470424, N470453, N470459, N470919, N470923, N470929, N470931, N470931, N470934, N470938, N470940, N470941, N470951, N470953, N470954, N471442, N471443, N471446, N471447, N471448, N471449, N471487, N471488, N471490, N471490, N471491, N471764, N471791, N471840, N471853, N471856, N471860, N471863, N471864, N471865, N471866, N471870, N471871, N471875, N471880, N472310, N472326, N472333, N472336, N472342, N472345, N472347, N472857, N472861, N472866, N472868, N472873, N472875, N472876, N472878, N472882, N472885, N472886, N472890, N472894, N472900, N472905, N473140, N473144, N473148, N473154, N473156, N473176, N473178, N473180, N473202, N473207, N473211, N473221, N473222, N473223, N473228, N473230, N473311, N473311, N473323, N473377, N473401, N473426, N473438, N473492, N473493, N473989, N473996, N473999, N474001, N474005, N474006, N474007, N474010, N474018, N474019, N474023, N474027, N474037, N474038, N474047, N474047, N474421, N474423, N474427, N474431, N474433.
Distribution pattern
Worldwide Distribution - USA (nationwide) including the states of AL,AR,AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, Ml, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WV., and the countries of Austria, Belgium, Cyprus, Denmark, France, Germany, Iraq, Italy, Lebanon, Netherlands, Panama, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, United Kingdom and Uruguay.