Event summary

Timeline bucket
July 21, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medtronic MiniMed Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combination with the Minilink transmitter and Sof-sensor glucose sensor for continuous monitoring of subcutaneous glucose concentration. The information provided by these devices is intended for use in identifying potential episodes of high or low blood glucose. The information provided is not intended to be used directly for making therapy adjustment but rather to assist the user in determining that an additional fingerstick glucose measurement may be appropriate. Per the product labeling, all therapy adjustments should be based on fingerstick glucose measurements obtained using standard home glucose meters and not on the sensor glucose values provided by the Guardian REAL-Time monitor. The CSS7100 and CSS7100K monitors are identical expect for the fact that the CSS7100K monitor is a pediatric version that has a minimum programmable low alert setting of 90 mg/dL. The minimum programmable low glucose alert setting for the adult model (CSS7100) is 40 mg/dL.

Z-2179-2014
Recall number
Z-2179-2014
Initiated
July 21, 2014
Classification
Class II
Status
Terminated
Recalling firm
Medtronic MiniMed Inc.
Quantity
86 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Medtronic MiniMed is recalling the Guardian REAL-Time monitors because of a manufacturing issue that has the potential to compromise the water resistance of the affected Guardian monitors.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Medtronic MiniMed is recalling the Guardian REAL-Time monitors because of a manufacturing issue that has the potential to compromise the water resistance of the affected Guardian monitors.

Code information

CSS7100 and CSS7100K

Distribution pattern

Worldwide:Distribution - US Nationwide and the countries of Croatia, Denmark, Finland, Germany, Netherlands, Norway, Sweden, Switzerland.