CloverSnare 4-Loop Vascular Retrieval Snare. Product is packaged in a Tyvek-film sterilizable outer package and is supplied one pouch in a box. The CloverSnare 4-Loop Vascular Retriever is intended for use in the cardiovascular system to manipulate and retrieve foreign objects, including but not limited to, wire guides, coils, balloons, catheters, and filters.
- Recall number
- Z-2243-2014
- Initiated
- July 17, 2014
- Classification
- Class I
- Status
- Terminated
- Recalling firm
- Cook Inc.
- Quantity
- 696 devices total distribution (671 nationwide)
App-derived interpretation
This remedial action is the result of six product complaints associated with separation of the snare from the distal tip of the wire.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
This remedial action is the result of six product complaints associated with separation of the snare from the distal tip of the wire.
Code information
Model Number(s): VRS-6.0-9.0 Lot numbers: 3583416, 3583418, 3583422, 3583424, 3583426, 3583428, 3583430, 3583432, 3583434, 3583436, 3583440, 3583442, 3583452, 3583456, 3583458, 3583462, 3583464, 3583466, 3583468, 3583470, 3583472, 3583474, 3583476, 3583478, 3583480, 3583482, 3583484, 3583486, 3583488, 3583490, 3583492, 3583494, 3583496, 3583498, 3583500, 3583502, 3583504, 4293921, 4293923, 4293925, 4293927, 4319573, 4319575, 4319577, 4319579, 4319581, 4319583, 4319585, 4319587, 4319589, 4319591, 4572365, 3583418X, 3583430X, 3583442X, 3583442XX, 3583464XXX, 3583480XX, and 3583486X.
Distribution pattern
Worldwide Distribution - USA including AK, AL, AZ, CA, CT, DC, FL, GA, HI, IA, ID, IL, IN, KY, LA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, PA, SC, TX, UT, VA, WA, WI, WV and Internationally to: Canada, Austria, Belgium, Denmark, Germany, Great Britain, Ireland, Italy, Spain, Sweden, and Switzerland.