Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69039

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 09, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Zimmer, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Persona The Personalized Knee System 2.5 mm Female Screw 48 mm Length Sterile Qty-2. Sterilized using irradiation, single use.

Z-2737-2014
Recall number
Z-2737-2014
Initiated
September 09, 2014
Classification
Class II
Status
Terminated
Recalling firm
Zimmer, Inc.
Quantity
16,380 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The Persona 48mm X 2.5mm Female Screw is being recalled due to potential use of the screw in cortical bone, and corresponding stripping of the 2.5mm Persona Hex Driver.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The Persona 48mm X 2.5mm Female Screw is being recalled due to potential use of the screw in cortical bone, and corresponding stripping of the 2.5mm Persona Hex Driver.

Code information

Lot Numbers: 62155015 62165014 62165015 62165016 62165017 62165018 62182742 62182743 62188282 62188283 62191355 62191361 62191366 62191372 62210290 62210291 62227943 62230668 62230669 62230670 62230671 62230672 62230673 62230674 62230675 62230676 62255591 62255592 62255593 62255594 62255595 62255596 62255597 62255598 62283576 62283577 62283578 62288220 62288221 62288222 62288223 62288224 62305911 62305912 62305913 62305914 62327983 62327985 62327987 62327988 62327989 62348078 62348079 62348080 62348091 62348092 62348095 62379888 62379889 62379890 62379891 62391801 62391802 62391803 62391804 62391805 62391807 62391808 62408432 62408438 62408439 62408440 62408441 62417035 62417036 62417037 62444920 62444921 62444922 62444925 62453583 62453584 62453585 62475432 62475433 62475434 62475435

Distribution pattern

Nationwide Distribution including the states of : HI, CA, UT, NY, OH, IN, AL, FL, GA, WI, IA, MN, NE, SD, NJ, PA, ND, IL, MO, VA, MD, NC, MI, OR, WA, KS, KY, TN, WV, OK, CT, MA, ME, AZ, NM, TX, LA, ID, CO, MT, UT, NV and AK., and Internationally to Singapore, Japan, Germany and Canada.