device · product 1 of 1
Coulter DxH Diluent, Catalog No. 628017 Blood cell analysis comprises diluting a whole-blood sample with a solution that functions as a diluent. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types and the measurement of hemoglobin.
Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.