device · product 1 of 1
TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.
- Recall number
- Z-2657-2014
- Initiated
- August 28, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Motion Concepts
- Quantity
- US: 59 units
App-derived interpretation
Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).
Code information
All units manufactured between June 2011 and May 2014: Serial Numbers 28583, 28597, 28604, 28607, 28629, 30334, 028380, U001050, U001083, U001544, U002020, U002326, U002386, U002403, U002411, U002489, U002502, U002505, U002509, U002510, U002519, U002716, U002922, U003577, U004011, U004018, U004073, U004079, U004145, U004214, U004400, U004404, U004472, U004485, U006042, U006234, U006235, U006260, U006295, U006296, U006353, U006378, U007126, U007138, U007163, U007239, U007245, U007279, U007288, U007407, U008556, U008631, U008698, U008840, U008896, U009548, U009569, U009579, U009707, U009718.
Distribution pattern
US Distribution in states of: AR, CA, CT, DE, IA, ID, IL, IN, MI, MD, MO, NE, NY, OH, PA, TN, TX, and WA.