Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69149

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 28, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Healthcare Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Dimension Vista¿ Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnostic product for the calibration of Creatine Kinase MB Isoenzyme (MMB) method on the Dimension Vista¿ System.

Z-0024-2015
Recall number
Z-0024-2015
Initiated
August 28, 2014
Classification
Class II
Status
Terminated
Quantity
6562

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf life of this product. Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been d

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf life of this product. Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been d

Code information

lot 3JD045, exp. 2014-10-01; lot 4BD064, exp. 2015-02-01; lot 4CD012, exp. 2015-03-01; and lot 4DD065, exp. 2015-05-01.

Distribution pattern

Worldwided distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Canada, Denmark, Germany, Finland, France, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, Slovenia, South Korea, Spain, and Switzerland.