Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69271

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 15, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
DePuy Orthopaedics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ATTUNE Intuition Distal Femoral Jig; sold non-sterile; product is wrapped in bubble wrap and then sealed inside a polyethylene bag

Z-0199-2015
Recall number
Z-0199-2015
Initiated
October 15, 2014
Classification
Class II
Status
Terminated
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
3,082

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Issuing a device correction because if the pin bushing is over loaded or inverted, there is the potential for partial or full displacement and fracture of the plastic around the pin bushing.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Issuing a device correction because if the pin bushing is over loaded or inverted, there is the potential for partial or full displacement and fracture of the plastic around the pin bushing.

Code information

Catalog NOs. 2544-00-521 and 2544-00-520; All lots are affected for the Device Correction. Lot numbers below should be inspected and swapped out if the bushing is incorrectly assembled: Part No. DePuy Synthes Lot No: MFG Lot No: 254400520 ABC22844 C22844 ABC22845 C22845 ABC34279 C34279 ABC34280 C34280 ABC4107 C4107 ABC4108 C4108 ABC4112 C4112 ABC4113 C4113 ABC50741 C50471 254400521 ABB13662 B13662 ABB19116 B19116 ABB19117 B19117 ABB19118 B19118 ABB35392 B35392 ABB35393 B35393 ABB39835 B39835 ABB41384 B41384 ABB46872 B46872 ABB49200 B49200 ABB50724 B50724 ABB55749 B55749 ABB63598 B63598 ABB66346 B66346 ABB72107 B72107 ABB77183 B77183 ABB79390 B79390 ABB79391 B79391 ABB80712 N80712 ABC17165 C17165 ABC5023 C5023

Distribution pattern

Worldwide Distribution - USA (nationwide) except AK, including Puerto Rico and the countries of Canada, Australia, Austria, Belgium, China, France, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Singapore, South Africa, South Korea, Spain, Switzerland, Thailand, UAE, United Kingdom and Chile.