Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69354

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 07, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Zimmer, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the Persona Tibial Articular Inserter is used to seat the articular surface implant onto the tibial baseplate.

Z-0123-2015
Recall number
Z-0123-2015
Initiated
October 07, 2014
Classification
Class II
Status
Terminated
Recalling firm
Zimmer, Inc.
Quantity
618 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Affected product did not undergo heat treatment hardening step leading to the potential for deformation of the tip.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Affected product did not undergo heat treatment hardening step leading to the potential for deformation of the tip.

Code information

Part Number 42-5299-001-00 Lots: 56572679; 56573170; 56572918; 56573378; 56572975; 56573379; 56572978; 56574223; 56572981; 56574291; 56572982; 56574292; 56573071; 56574316; 56573115; 56574317; 56573116; 56574318; 56573154.

Distribution pattern

Worldwide Distribution - USA (nationwide) including the states of NY, OR, NC, OH, TX, FL, CO, IN, TN, PA, CA, KS, MO, WI, VA, MN, NJ, UT, MA, GA, MI, OK, AZ, MS, IL and AL., and the countries of Korea, Singapore, Canada and Japan