Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69560

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 21, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Maquet Medical Systems USA

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

TEGRIS System manufactured by MAQUET GMBH in Germany The Maquet Tegris OR Integration System is designed to be used as the central operating unit in an operating room. The integration system has two main functions: Recording and distribution of images and videos and interaction with medical and non-medical devices.

Z-0993-2015
Recall number
Z-0993-2015
Initiated
August 21, 2014
Classification
Class II
Status
Terminated
Quantity
6 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design (manufacturing process)

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.

Code information

part number: 1703.35AO, with software version up to 2.1.x.x.

Distribution pattern

US Nationwide Distribution including NE, NY and PA.