Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69603

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 23, 2014
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Synthes (USA) Products LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System. Intended for posterior pedicle screw fixation (Tl OS2- ilium), posterior hook fixation (Tl-L5), or anterolateral fixation (T8-L5).

Z-0226-2015
Recall number
Z-0226-2015
Initiated
October 23, 2014
Classification
Class I
Status
Terminated
Quantity
24

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct specification.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct specification.

Code information

Part Number 04.634.002 Lot Number 6816781

Distribution pattern

Distributed to MI, MT, and CO.