Event summary

Timeline bucket
November 06, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Fisher & Paykel Healthcare, Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Fisher and Paykel Healthcare (FPH) IW900-Series Infant Warmer Fisher and Paykel; Healthcare CosyCot and Kit Replacement Head Case Assembly It is designed to assist with the thermoregulation of infants weighing up to 10kg and is used by labor and delivery rooms and pediatric intensive care units. Infant Warmers Model Number (REF) IW910XXX, IW920XXX, IW931XXX, IW932XXX, IW933XXX, IW934XXX, IW951XXX, IW952XXX, IW953XXX, IW954XXX, IW980XXX, IW990XXX Spare Parts Model Number (REF) 043041129, 043041130, 043041131, 043042359, 648040142

Z-0210-2015
Recall number
Z-0210-2015
Initiated
November 06, 2014
Classification
Class II
Status
Terminated
Quantity
Nationwide (U.S.): 661 Infant Warmers and 180 Spare Parts, Internationally: 2,451 Infant Warmers and 280 Spare Parts

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Manufacturing material removal

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.

Code information

Affected Serial (SN) and Lot (LOT) Numbers Infant Warmers: 071116ZZZZZZ - 090810ZZZZZZ Spare Parts: 071116 - 090810

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Bekgium, Bermuda, Chile, China, Croatia, Dominican Republic, Estonia, Finland, France, Germany, Great Britain, Greece, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Mexico. Morocco, New Zealand, Nigeria, Northern Ireland, Norway, Pakistan, Palestinian Territory, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, singapore, Slovenia, South Africa, Spain, Sri Lanka, Sultanate of Oman, Syria, Taiwan, Thailand, The Netherlands, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam, and West Bank.