device · product 1 of 1
Breas Vivo 50 Home Care Ventilator Product Usage: Respiratory Care
- Recall number
- Z-2094-2015
- Initiated
- November 07, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Breas Medical AB
- Quantity
- 846 US
App-derived interpretation
Unintended treatment termination could result from a keypad malfunction in some situations. The device erroneously interprets this as a Stop Treatment Instruction. An alarm will not sound, or be registered. Accessories and monitoring equipment connected to the Vivo 50 will stop functioning as the device enters stand-by mode.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Unintended treatment termination could result from a keypad malfunction in some situations. The device erroneously interprets this as a Stop Treatment Instruction. An alarm will not sound, or be registered. Accessories and monitoring equipment connected to the Vivo 50 will stop functioning as the device enters stand-by mode.
Code information
All Vivo 50 Ventilators (215000, 215016) with firmware version 2.04
Distribution pattern
Worldwide - United States Nationwide Distribution and the countries: Argentina, Belgium, Chile, Cyprus, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Malaysia, Netherlands, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Thailand, Turkey, UAE and United Kingdom.